IiNkqubo zokuJonga izigulane zeMfuyo HD10

HD-10


Iinkcukacha zeMveliso

Imodyuli yeECG

Imodyuli yokuphefumla

Ifuthe legazi

Imodyuli ye-Oximetry

Ubushushu bomzimba

Umoya omdaka

Ubume bomgangatho: IECG, ioksijini yegazi, uxinzelelo lwegazi, izinga lokuphefumla, ubushushu, izinga lentliziyo.
Ukucwangciswa okukhethiweyo: isikrini se-capacitive, umshicileli, i-WIFI, i-mainstream / bypass ETCO2, i-Masimo yegazi, i-Suntech blood pressure, njl.
Izinto ezincedisayo: i-five-lead heart wire * 1, blood oxygen probe * 1, blood pressure extension tube * 1, blood pressure cuff * 4, disposalable electrode sheet * 15, power cord * 1.
Izixhobo ezikhethiweyo: iidoli, izibiyeli zodonga, njl.

Inkqubo yokusebenza yeLinux, esekelwe kubasebenzisi abaninzi, imisebenzi emininzi, inkqubo yokusebenza ye-CPU emininzi;ujongano oluhle lokusebenzisana kunye nokuhamba okukhuselekileyo nozinzo lokusebenza;HD10VETesweni izilwanyanaixhotyiswe ngenkqubo yokusebenza yelinux ukuphumeza ukwandiswa kwethekhnoloji ye-multi-domain;Uqhagamshelo lwe-WIFI kwisikhululo sokubeka iliso esisembindini, kunye nokuphuculwa kwe-USB, ukwandiswa kweTV kunye nolunye ukubekwa kweliso kwesikrini esikhulu (ukhetho), njl.

Indlela yokubonisa isixhenxe-ikhokelela kwisikrini esifanayo.
Uluhlu lokulinganisa izinga lentliziyo: 20-350bpm.
Uluhlu lwe-alam: 20-350bpm.
Umlinganiselo wokulahlwa kwendlela eqhelekileyo ≥105db.
Isantya seWaveform 6.25, 12.5, 25, 50mm/s ezine amanqanaba okukhetha.
Fumana ukhetho Iindlela ezisixhenxe zokukhetha.
Umthamo we-QRS onamanqanaba ama-5 ohlengahlengiso.
Imowudi yokubeka iliso Ukuxilongwa, ukubeka iliso, utyando, ST iindlela ezine zokubeka iliso.
Uhlalutyo lwezinga lentliziyo ye-arrhythmia Ngeentlobo ezingama-23 zohlalutyo lwezinga lentliziyo ye-arrhythmia.
Ukukhuselwa kwe-Electrodebrider Imodyuli ye-ECG inokusebenza ngokuqhelekileyo kwindawo yokuphazamiseka kwe-electrosurgical kunye namandla okusika i-electrodebrider ye-300W kunye namandla okudibanisa i-100W, kwaye ukuchaneka akuthinteki.

Uluhlu lokulinganisa: 0 - 150 ukuphefumula / min.
Ukuphefumla okuvuselelweyo Umqondiso we-Sine wave, 62.8kHz (±10%), <500μA.
I-alam ye-asphyxiation enamanqanaba angaphezu kwe-9 ye-alamu yokumisela umda wexesha.
Ukufumana uhlengahlengiso kunye namanqanaba e-8 ohlengahlengiso lokuzuza.

Uluhlu lwemilinganiselo yoxinzelelo lwegazi (iiyunithi) i-Equine Canine Feline
Uxinzelelo lwegazi lweSystolic mmHg 40-255 40-200 40-135
kPa 5.3-34.0 5.3-26.7 5.3-18.0
Uxinzelelo lwegazi lweDiastolic mmHg 20-205 20-150 20-100
kPa 2.7-27.3 2.7-20.0 2.7-13.3
Uxinzelelo oluphakathi mmHg 27-220 27-165 27-110
kPa 3.6-29.3 3.6-22.0 3.6-14.7
Uluhlu lomlinganiselo wecuff: 0 ~ 290±3mmHg.
Ukhuseleko loxinzelelo olugqithisileyo lweSoftware Equine: hayi ngaphezulu kwe-293 mmHg
Canidae: akukho ngaphezulu kwe-250 mmHg.
I-Feline: ayidluli kwi-148 mmHg.
Ukhuseleko lwe-Hardware overpressure Equidae/canines: ukuya kuthi ga kwi-300 mmHg.
Inyama: ukuya kuthi ga kwi-150 mmHg.
Ixesha lokulinganisa kwimodi ezenzekelayo 2.5 min, 5 min, 10 min, 15 min, 20 min, 30 min, 45 min, 60 min, 90 min, 120 min.

 

Uluhlu lokulinganisa: 69% -100%.
Ithowuni yePulse Ngomsebenzi wokutshintsha ithowuni ye-pulse.
Isantya se-Waveform 3 okanye ngaphezulu iinketho zesantya sohlobo lwamaza.
Uvakalelo Iinketho ezi-3 zobuntununtunu ezikhoyo.

I-HD10Vet esweni isilwanyana yamkela ukucwangciswa kwemodyuli yobushushu bomzimba kabini, kunye ne-channel-channel umehluko weqondo lokushisa.
umboniso wexesha lokwenyani wexabiso elipheleleyo lomahluko wobushushu obulinganisiweyo.

Uluhlu lokulinganisa: 0mmHg-114mmHg
Umsebenzi wembuyekezo kunye ne-O2, imbuyekezo ezenzekelayo yoxinzelelo lwe-atmospheric, imbuyekezo ye-N2O
Umsebenzi wokulinganisa: FiCO2, EtCO2, awRR