Ukucushwa okujwayelekile: I-ECG, umoya-mpilo wegazi, umfutho wegazi, izinga lokuphefumula, izinga lokushisa, ukushaya kwenhliziyo.
Ukucushwa ongakukhetha: isikrini esinamandla, iphrinta, i-WIFI, i-mainstream/bypass ETCO2, i-Masimo blood oxygen, i-Suntech blood pressure, njll.
Izesekeli: izintambo zenhliziyo ezine-lead emihlanu * 1, i-blood oxygen probe * 1, ishubhu yokwandisa umfutho wegazi * 1, i-blood pressure cuff * 4, ishidi le-electrode elahlwayo * 15, intambo yamandla * 1.
Izesekeli ozikhethela zona: onodoli, abakaki bodonga, njll.
Isistimu yokusebenza ye-Linux, esekelwe kubasebenzisi abaningi, i-multi-tasking, isistimu yokusebenza yama-CPU amaningi;isikhombimsebenzisi esihle sokusebenzisana kanye nokugeleza kokusebenza okuphephile nokuzinzile;Idatha ye-HD10VETukuqapha isilwaneifakwe uhlelo lokusebenza lwe-linux ukuze kuzuzwe ukunwetshwa kobuchwepheshe besizinda esiningi;Ukuxhumeka kwe-WIFI esiteshini esimaphakathi sokuqapha, kanye nokuthuthukiswa kwe-USB, ukunwetshwa kwe-TV nokunye ukuqapha kwesikrini okukhudlwana (uma uthanda), njll.
Indlela yokubonisa eziyisikhombisa-lead esikrinini esifanayo.
Ibanga lesilinganiso senhliziyo: 20-350bpm.
Ibanga le-alamu: 20-350bpm.
Isilinganiso sokwenqatshwa kwemodi evamile ≥105db.
Isivinini se-Waveform 6.25, 12.5, 25, 50mm/s amazinga amane okukhetha.
Thola ukukhetha Izindlela eziyisikhombisa zokukhetha.
Ivolumu ye-QRS enamazinga angu-5 okulungisa.
Imodi yokuqapha Ukuxilonga, ukuqapha, ukuhlinzwa, ST izindlela zokuqapha ezine.
Ukuhlaziya izinga lokushaya kwenhliziyo nge-arrhythmia Ngezinhlobo ezingama-23 zokuhlaziywa kwe-arrhythmia yenhliziyo.
Ukuvikelwa kwe-Electrodebrider Imojula ye-ECG ingasebenza ngokujwayelekile endaweni yokuphazanyiswa kwe-electrosurgical enamandla okusika i-electrodebrider angu-300W namandla okuhlanganisa angu-100W, futhi ukunemba akuthinti.
Ibanga lokulinganisa: 0 – 150 ukuphefumula/ngomzuzu.
Ifomu legagasi elishukumisa ukuphefumula Isignali yegagasi ye-Sine, 62.8kHz (±10%), <500μA.
I-alamu yokuphefumula enamazinga angaphezu kwangu-9 okulungiselelwa komkhawulo we-alamu.
Thola ukulungiswa ngamaleveli angu-8 okulungiswa kwenzuzo.
Ibanga lokulinganisa Umfutho wegazi (amayunithi) I-Equine Canine Feline
Umfutho wegazi we-Systolic mmHg 40-255 40-200 40-135
kPa 5.3-34.0 5.3-26.7 5.3-18.0
Umfutho wegazi we-Diastolic mmHg 20-205 20-150 20-100
kPa 2.7-27.3 2.7-20.0 2.7-13.3
Ingcindezi emaphakathi mmHg 27-220 27-165 27-110
kPa 3.6-29.3 3.6-22.0 3.6-14.7
Ibanga lokulinganisa ingcindezi ye-Cuff: 0 ~ 290±3mmHg.
Ukuvikelwa kokucindezela ngokweqile kwesofthiwe I-Equine: akukho ngaphezu kuka-293 mmHg
I-Canidae: akukho ngaphezu kuka-250 mmHg.
I-Feline: akufanele idlule ku-148 mmHg.
Izingxenyekazi zekhompuyutha zokuvikela ukucindezela ngokweqile kwe-Equidae/canines: kufika ku-300 mmHg.
Inyama: kufika ku-150 mmHg.
Isikhawu sokulinganisa kwimodi ezenzakalelayo 2.5 min, 5 min, 10 min, 15 min, 20 min, 30 min, 45 min, 60 min, 90 min, 120 min.
Ibanga lokulinganisa: 69% -100%.
Ithoni ye-Pulse Ngomsebenzi wokushintsha ithoni ye-pulse.
Isivinini se-Waveform 3 noma izinketho eziningi zesivinini se-waveform.
Izinketho zokuzwela 3 ziyatholakala.
I-HD10Vet yokuqapha isilwane isebenzisa ukucushwa kwemojula yezinga lokushisa lomzimba elikabili, ngokubonisa umehluko wezinga lokushisa elikabili-iziteshi ezimbili.
isibonisi sesikhathi sangempela senani eliphelele lomehluko wezinga lokushisa olinganisiwe.
Ibanga lokulinganisa: 0mmHg-114mmHg
Umsebenzi wokunxeshezelwa nge-O2, isinxephezelo esizenzakalelayo sokucindezela komkhathi, isinxephezelo se-N2O
Umsebenzi wokulinganisa: FiCO2, EtCO2, awRR